Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Astro-Med MT95K2 Gain (1 year) CAL VER /5520 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 39 CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Astro-Med MT95K2 Verification Checklist # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Connect the 5520A NORMAL HI and LO to UUT [MEM2]. 1.002 RSLT =[MEM2] 1.003 MESS Slew output for a full scale deflection (positive). 1.004 5520 250mV 0.8% 2W 2.001 MESS Slew output for a full scale deflection (negative). 2.002 5520 -250mV 0.8% 2W 3.001 END